PRELIMINARY OPERATIONS

-clean the surfaces to be measured

-clean the feeler gauge, in particular the sphere and the measuring tip

- bring the sphere closer to the tip so that it touches the flat part with the sensor (zero value)

- hold the tip with the ball in a vertical position

- switch on the thickness gauge by connecting the power cable

-Wait for the end of the initial calibration

- Carry out some measurements with no load  you will notice a flashing W indicating Wait and after a few flashes an @ will appear indicating that it is  calibrated.

-For a more accurate measurement, work at room temperature, carry out some measurements with no load on a thickness of ~ 10 mm and wait at least 5 minutes.

​

​

MEASUREMENT

-Place the sphere on the surface  to be measured, while on the opposite surface in correspondence with the sphere, place the tip

-by magnetism the sphere is attracted to the tip without the need for forcing

- read the measured quota

- extract the sphere from the piece under examination

-Turn off the SpeXor by unplugging the power cord.

-After taking the measurement and without moving the tip, it is possible to reset (0.00) the measured value by pressing the button and the subsequent measurement will detect the difference between the first and second measurement (positive or negative with sign -  before the value).

​

​

     ATTENTION! the magnetic attraction force between the tip and the sphere may not be sufficient to prevent the latter from falling, PAY ATTENTION!

​

resolution.png

The table is purely indicative and does not represent a guarantee. The measurements were taken at an ambient temperature of 20°C.

The sphere, if used improperly, can fall. Measurements above 13 mm must be taken from below, as shown in the image. However, measurements taken differently than the image will be affected by gravity.

Components.JPG

Unit of measure                 millimeters / inches (on request)

-------------------------------------------------- -

Measurement system         Hall effect sensor

-------------------------------------------------- ---

Display                                  Liquid crystal digital LCD

-------------------------------------------------- ---

Alimentazione                      5V USB cable

-------------------------------------------------- ---

Usage temperature               + 5 ° ~ + 40 °

1 Magnetic sphere Ø10mm

2 Tip with hall effect sensor

3 Handle

4 Neoprene cable

5 LCD display

6 ZERO reset button

7 USB port

The Spexor is very precise in the range from 0 to 5 mm where it is possible to stay in the tenth of a mm with appropriate precautions,  to use it as a measurement tool make sure to let it self-calibrate before each measurement, see on the MORE page the video showing an @ appearing when recalibrating. The temperature should be room if possible.

In any case, the complicated self-calibration algorithm will allow you to detect your Hollow form with extreme repeatability, not for nothing that the turners who do piercing and carving consider it an indispensable tool.

WARRANTY: 24 MONTHS